Insulation Resistance:
- Test voltage 250V/500V/1000V
- Measuring ranges (Mid-scale value) 20MΩ/200MΩ/2000MΩ
- Output short circuit current 1.5 mA DC approx.
- Accuracy: ±1.5%rdg±5dgt (20MΩ/200MΩ), ±10%rdg±3dgt (2000MΩ)
Continuity Test
- Measuring ranges: (Mid-scale value) 20Ω/200Ω/2000Ω
- Output voltage on open circuit 7 - 12V DC
- Measuring current 200mA DC min.
- Accuracy: ±1.5%rdg±5dgt (20Ω), ±1.5%rdg±3dgt (200Ω/2000Ω)
AC Voltage
- AC voltage range 0 - 600V AC, ±5%rdg±3dgt
- Safety Standard IEC 61010-1 CAT III 300V Pollution Degree 2 IEC 61010-2-031 IEC 61557-1/2/4
- IP Rating IEC 60529-IP54
- EMC Standard IEC 61326-1
- Power source R6 (AA) (1.5V) × 8
- Dimensions 167(L) × 185(W) × 89(D)mm
- Weight 970g approx
Warranty:
Brand Warranty
When installing new electrical machinery or equipment, testing insulation resistance is important for two reasons. First, it ensures that the insulation is in adequate condition to begin operation. ... The test is accomplished by applying DC voltage through the de-energized circuit using an insulation tester.